Technical Speci cations
Measurements procedures:
• Endpoint, Fixed Time, Kinetic, Bichromatic
• Blank: Water, Reagent blank or Sample blank
• Calibration: Factor, Calibrator and Multipoint
• Up to three di erent reagent additions possible for
each method
• Samples can be loaded by single test or pro le
• 9 di erent calculation algorithms programmed for lin-
ear and non-linear multipoint analysis
Operation:
• Random access 130 tests/hr
• Runs clinical chemistry and turbidimetric tests
• Large choice of Cypress Diagnostics reagents available
• Up to 300 clinical applications programmable
Photometer particulars:
• Multi wavelength double beam photometer with nar-
row band interference lters
• Filters (8): 340, 380, 405, 492, 510, 546, 576, 620 nm
• Halogen lamp 6V/10W (2000 hours)
• Linear range: 0,0005 – 3,5000 ABS
• Resolution: 0,0005 ABS
• Drift: < 0,0001 ABS
• Detector: Silicon Diode
• Automatic autozeroing on the sample and reference
channel (double beam)
Option:
• Barcode reader for reagents and samples
Sampling:
• High precision diluter module
• Sample aspiration: 2-99 µl with 1 µl increments
• Reagent 1 volume: 3-500 µl with 1 µl increments
• Reagent 2 & 3 volumes: 3-330 µl with 1 µl increments
• Minimum reaction volume: 220 µl
• Maximum reaction volume: 550 µl
• Precision: CV < 1% at 2 µl
• Liquid level sensor
• Preheated sampling line at 37°C
• Automatic probe wash cycle in between tests to avoid
cross-contamination
Cuvettes:
• 120 semi-disposable reading cuvettes (6 racks with 20
positions) with a 6 mm light path
• Automatic washing and drying cycle by the washing
station for reusing the cuvettes
• Automatic QC of cuvettes after washing
• Temperature controlled at 37°C
Capacity:
• 32 sample positions (4 racks with 8 positions) of 3 ml
sample cups or primary tubes
• 24 reagent containers (35 ml or 6 ml)
• 12 positions for controls and calibrators
Operating environment:
• In covered and dry places
• Temperature: 18°C – 32°C
• Relative humidity: 20 %-85 %
• Stable surface (no vibrations)
• > 10 cm around the instrument to permit air circulation
Weight & dimensions:
• 53 x 57 x 75 cm (H x W x L)
• 37 Kg
Power:
• 110/240 V – 47/63 Hz
• 300 W
• Advised:
Electronic Stabilizer (> 1kW Power potential)
No-Break Module (UPS - Uninterrupted Power Supply)
Software features:
• Unlimited work list
• Historical patient results database
• Methods, calibrators and controls database
• Method programming includes control parameters
• STAT-samples
• Quality Control and visualization and printout of graphs
(Levey-Jennings)
• Sample pre-dilution and automatic dilution in rerun
mode
• Automatic rerun programmable
• Reference values are user de nable for serum and urine
for child, female and male
• Sample entry for serum/urine
• Multipoint calibration curve (automatic dilution of mas-
ter standard) and relative graphs
• Bidirectional Interface with Host Computer
• Auto-Diagnostics Program for a system built on di er-
ent modules
• Communication: USB 1.1/2.0 Full Speed & RS-232
• Operative System: Windows XP, Vista & Windows 7
• Language: English, French and other languages upon
request